6.10.2 Excursion without validation

4. When an excursion occurs, an excursion sample captured dialog box opens, and a
validation occurs. See
Following the excursion and validation, a validation results dialog box opens that
contains the blank and 500 ppb sucrose data. The next arrow on the validation
results dialog box takes you to the validation summary dialog box.
The analyzer writes the date, time, TOC value, analyzer serial number, status,
oxidation curve filename, conductivity, and temperature information to the RFID
tag.
Figure 95 Excursion sample captured dialog box
The excursion sample dialog box displays the analyzer's serial number, the date and time
of the excursion, the TOC value, and the oxidation curve filename.
Export icon opens the export sample dialog box. See
export the oxidation curve file to a USB flash drive.
Yes icon (containing checkmark) returns you to the load bottles dialog box.
Cancel icon takes you to the unload/replace bottles dialog box.
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1. Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See
Figure
96. From left:
Bottle 1 = empty
Bottle 2 = empty
Bottle 3 = empty
Bottle 4 = excursion bottle
Figure
95.
Bottle Mode
Figure
90. From there, you can
Page 113