2.1.3 Analysis Data Screen

B I O T E C T O R
0 9 : 1 3 : 0 2
1 2 - 0 9 - 0 2
T I C & T O C
R E A C T I O N
R E S U L T
0 9 : 0 7 : 0 2
1 2 - 0 9 - 0 2
0 9 : 0 1 : 0 2
1 2 - 0 9 - 0 2
0 8 : 5 5 : 0 2
1 2 - 0 9 - 0 2
0 8 : 4 9 : 0 2
1 2 - 0 9 - 0 2
0 8 : 4 3 : 0 2
1 2 - 0 9 - 0 2
0 8 : 3 7 : 0 2
1 2 - 0 9 - 0 2
The Analysis Data screen is the default display screen on the BioTector for carbon analyses (TIC, TOC, TC,
VOC) in mgC/l, COD & BOD results in mgO/l, and Drinking Water parameters, which are REMOVAL in %,
and RESULT in FAIL/PASS condition depending on the analysis type and specific configuration settings.
When the user moves through the various levels of the software menu, BioTector returns to this screen
automatically after 15 minutes if there is no further activity on the membrane keypad.
This screen gives information on:
- The Reaction Start time.
- The Reaction Type, for example a TIC & TOC reaction, TC reaction, Cleaning Reaction.
- The Reaction Phase, for example if the reaction is currently in the TIC, Base Oxidation, TOC phase.
- The operation Range (e.g. Range 1, 2 or 3) the BioTector is using to carry out its analysis.
- The Reaction Time, which is the elapsed time (seconds) since the analysis start.
- The Reaction Duration, which is the overall duration (seconds) of the analysis.
The Analysis Data screen also has an archive of the last 25 reactions. The most recent six reactions are
shown on the screen. In order to access the remaining reactions, use the DOWN or RIGHT keys to scroll
down, use the LEFT or UP keys to scroll up.
Each reaction record in the reaction archive contains:
- Start Time - reaction start time.
- Date - reaction date.
- Record Type, using the prefixes below:
– reactions from stream 1 to stream 3.
S1 to S3
– reactions from manual sample stream 1 to manual stream 3.
M1 to M3
– sample sensor detected the sample or there is no significant quantity of air bubbles in the
stream/manual grab sample lines.
– sample sensor detected no sample or there is significant quantity of air bubbles in the
x
stream/manual grab sample lines. See Sample Status in section
details.
– remote standby reaction.
RS
– zero calibration reaction.
ZC
– zero check reaction.
ZK
– manual zero adjust.
ZM
– span calibration reaction.
SC
– span check reaction.
SK
– manual span factor adjust.
SM
– 24 hours average result from stream 1 to stream 3.
A1 to A3
- Analysis Results – analysis results according to the analysis type (e.g. TIC, TOC in mgC/l).
R U N N I N G
R E A C T I O N
S T R E A M 2
R E A C T I O N
T O C
R E A C T I O N
1
R A N G E
2 6 6 s
R E A C T I O N
3 6 0 s
R E A C T I O N
S 1 √
S 2 √
S 3 √
S 4 x
S 5 x
C F
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
S T A R T
T Y P E
P H A S E
T I M E
D U R A T I O N
T I C m g C / l
T O C m g C / l
1 3 0 . 0
3 . 6
7 . 2
1 0 . 7
1 4 . 3
0 . 9
5 4 0 . 0
3 . 6
7 . 2
1 0 . 7
1 4 . 3
7 . 9
8.3.8 Fault Setup