2.2.3 Mounting to dual instrument poles

Follow these guidelines and refer to
instrument poles:
Mount the analyzer mounting bracket to two instrument poles that are attached to a
common surface. Do not secure to two instrument poles that could move
independently.
The instrument poles should extend at least 10 inches (254 mm) from a common rigid
base and should be no larger than 2 inches (50.8 mm) in diameter.
Use four 5/16-inch U-bolts, two for each 2-inch pipe, and eight matching nuts, to
mount the analyzer mounting bracket to two rigid instrument poles. Use U-bolts and
nuts that can withstand the process environment. HACH Company does not supply
U-bolts or nuts.
The U-bolts and nuts should not extend more then 0.60 inches (15 mm) from the
inside of the mounting bracket to allow clearance for the analyzer.
Figure 5 PAT700 instrument-pole mount
Figure 2
and
Figure 5
to mount the analyzer to
Installation
Page 17