6.5.2 TOC calibration procedure

Bottle selection check boxes enable you to enable or disable the bottles to use in the
test. The default setting is all three bottles checked. At least one bottle must be selected.
Reps text box allows you to specify the number of repetitions to perform on each bottle.
You may enter a value from 1 to 3. The default setting is 3.
Save as default button saves the bottle selection and reps as the default setting.
Back icon returns you to the TOC calibration setup dialog box. All changes made to
custom setup are retained.
Forward icon takes you to the load bottles dialog box. All changes made to the custom
setup are retained. See
Figure
Cancel button takes you to the run standards dialog box. All changes made in any of the
dialog boxes are lost.
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1. Allow the TOC calibration standards to warm to room temperature.
2. At the home screen, touch the << icon to open the sliding toolbar, then touch the
bottles icon in the toolbar, or touch the bottles icon in the lower left corner of the
screen.
3. Touch the run standards button in the bottle mode dialog box. See
page
88.
4. Touch the TOC calibration button in the run standards dialog box. See
page
90.