5.9.3 TOC validation tab
5.9.4 System suitability tab
Figure 56 Bottle test setup dialog box, TOC validation tab
Return to online mode checkbox enables the analyzer to return to the online mode
following the validation. You may enable or disable (default) this option. When enabled,
the analyzer returns to online mode following the validation. When disabled, the analyzer
stays at the bottle mode dialog box. All valves are closed, the pump is off, and the lamp is
off.
Return icon returns you to the setup dialog box and saves any changes. Any changes
are noted in the audit trail.
Figure 57 Bottle test setup dialog box, system suitability tab
Return to online mode checkbox enables the analyzer to return to the online mode
following the system suitability test. You may enable or disable (default) this option.
When enabled, the analyzer returns to online mode following the system suitability. When
disabled, the analyzer stays at the bottle mode dialog box. All valves are closed, the
pump is off, and the lamp is off.
Return button returns you to the setup dialog box and saves any changes. Any changes
will be noted in the audit trail.
Setup
Page 79