Setup

5.8.3 Users tab

Page 72
Remember last X passwords text box allows you to set the number of passwords to
keep in memory to prevent a user from using an old password. You may enter a value
from 1 to 5 passwords. The default value is 3.
Automatically logoff after X minutes text box allows you to set the number of minutes
of inactivity before auto logoff. You may enter a value from 0 to 60 minutes. A setting of 0
disables auto logoff. The default value is 30.
Return icon returns you to the setup dialog box and saves the changes. Any changes
are noted in the audit trail.
The users tab allows you to add, edit, or delete users of the analyzer. Up to 10 users can
be configured for the analyzer. Administrator and factory users cannot be deleted.
Figure 49 Security settings dialog box, users tab
Four different user types may access the system: guest, operator, administrator, and
factory. Only operator users may be added or deleted from the system.
Guest. This user type is any user who is not logged on to the system. This user may
access most areas and view information but is not allowed to modify any settings or
initiate any operations. Guests have no access to the security settings dialog box.
Operator. This user type has full access to most dialog boxes. Operators have no
access to the security settings dialog box and read-only access to the factory settings
and the bottle mode settings.
Administrator. This user type has full access to most dialog boxes. Administrators
have read-only access to the factory dialog box. The default administrator entry has a
user name of "Administrator", a user ID of "Admin", and a password of "123456".
Factory. This user type has full access to all dialog boxes.