5.11.4 User slopes tab

5.11.5 Lamp tab

Figure 62 Factory settings dialog box, user slopes tab
TOC text box (read only) allows you to view the TOC calibration factor. This value is
updated after a TOC calibration has been performed.
Conductivity text box (read only) allows you to view the conductivity calibration factor.
This value is updated after a conductivity calibration has been performed.
Return icon returns you to the setup dialog box and saves the changes. Any changes
are noted in the audit trail.
Figure 63 Factory settings dialog box, lamp tab
Warning text box (read-only) allows you to view the factory-set value for the UV
Detect™ sensor output to be measured before a lamp is marginal.
Failure text box (read-only) allows you to read the factory-set value for the UV Detect
sensor output to be measured before a lamp has failed.
Return icon returns you to the setup dialog box.