Alarms

7.4.1 Error alarms

Code
Message
108
UnableToReadSettings
109
UnableToWritetoSettings
112
UnableToStartPrinterModule
113
UnableToStartRS232Module
115
UnableToStartLamp
116
UnableToReadDataLog
117
UnableToWriteToDataLog
123
UnableToReadAlarmLog
124
UnableToWritetoAlarmLog
128
UvLampMonitorFailed
140
UnableToWriteAuditTrailEntry
161
TocAlarm
162
ConductivityAlarm
163
UnableToStopCurrentOperation
164
UnknownError
167
WatchDogIOCard
168
WatchDogTocAlgoithm
169
WatchDogPrinter
170
WatchDogRs232
171
WatchDogModbus
177
Exception
Page 120
Errors occur any time one of the non-critical subsystems fail or a failure occurs from
which the analyzer attempts to recover. The animation area flashes red to indicate an
error condition is present. When an error condition is present, the analyzer attempts to
recover and continue operating. See
Table 14 Error alarm messages
Table
14.
Alarm condition
Some or all of the settings could not be read from the settings file.
Analyzer cannot write to settings file.
Analyzer cannot start the printer module.
Analyzer cannot start the RS-232 module.
Analyzer cannot start the lamp module.
Analyzer cannot read data log.
Analyzer cannot write to data log.
Analyzer cannot read alarm log.
Analyzer cannot write to alarm log.
UV lamp monitor failed.
Analyzer cannot write audit trail entry.
TOC value has exceeded the configured TOC alarm limit.
Conductivity reading exceeds the USP <645> conductivity limits.
Analyzer cannot stop its current operation.
Analyzer has experienced an unknown error.
I/O Card has stopped responding.
TOC algorithm has stopped responding.
Printer has stopped responding.
RS-232 module has stopped responding.
Modbus module has stopped responding.
An exception was thrown.