Advertisement

Setup

5.3.4 Stabilization tab

Figure 36 TOC dialog box, stabilization tab
When the analyzer switches to perform an on-line TOC analysis, after running the
analyzer in continuous flow modes such as idle with conductivity, conductivity mode or
standby for extend periods of time, a thermal stabilization and cleaning of the analysis
cell can reduce the number of TOC analysis cycles necessary to produce stable results.
An automatic stabilization process, along with its activation and duration criteria, can be
enabled in the stabilization dialog box. The stabilization process involves trapping water
in the cell and turning on the UV lamp to thermally stabilize and clean the cell.
Stabilize cell prior to analyzing check box allows you to enable automatic stabilization
of the cell prior to performing a reportable TOC analysis.
Time between TOC analyses text box allows you to enter the minimum elapsed time
between TOC analyses required to activate the automatic stabilization process. You may
specify a time in hh:mm:ss format between 00:00:01 and 23:59:59. The default value is
02:00:00.
Stabilization time text box allows you to enter the amount of time to run the stabilization
process. You may specify a time in hh:mm:ss format between 00:00:01 and 23:59:59.
The default value is 00:10:00
Page 59