Setup

5.5 Alarms

Page 64
The alarms icon allows you to enable or disable TOC and conductivity alarms for the
system.
TOC upper limit text box allows you to set the alarm upper limit for online TOC
measurement. You may enter a value from 0 to 2000. The default value is 500. The value
applies only to online TOC analysis, not to bottle tests or grab sample analysis.
TOC enabled checkbox allows you to enable or disable the online TOC alarm. If
enabled, any online TOC reading that exceeds the alarm upper limit triggers a TOC
alarm. If disabled, no TOC alarms are triggered. The default setting is enabled.
Conductivity enabled checkbox allows you to enable or disable the conductivity alarm.
If enabled, any conductivity reading that exceeds the value in
temperature triggers a conductivity alarm. If disabled, no conductivity alarms are
triggered. The default setting is enabled.
Alarms are reported for conductivity excursions only in the uncompensated mode and
require that uncompensated conductivity be selected.
Note: For proper operation of the conductivity alarm, uncompensated conductivity must be selected
under the general tab of the System Setting dialog box. Refer to
Figure 42 Alarm dialog box
Table 12
for the current
Display tab, page
61.