Bottle Mode

6.1.5 System suitability protocol

6.2 Accessing bottle mode

6.3 Bottle mode dialog box

Page 88
Perform a system suitability test according to the guidelines established in USP Method
<643> using prepackaged 500 ppb sucrose and 500 ppb 1,4-benzoquinone standards.
To access bottle mode, follow either of these procedures:
Touch the << icon to open the sliding toolbar, then touch the bottles icon in the toolbar,
or
Touch the bottles icon in the lower left corner of the home screen.
Note: If security is enabled, you must be signed on the analyzer to access the bottle mode. Bottle
tests can not be cancelled or accepted/rejected when security is enabled and a user is logged off the
analyzer.
The bottle mode dialog box allows you to perform procedures using the bottles in the
bottle bay. You may run calibrations, validations, or system suitability tests, analyze
samples from a bottle (grab sample), or install an excursion bottle to capture a sample
when a TOC measurement exceeds an alarm limit, a conductivity alarm occurs, specific
alarm conditions occurs, or manually. See
Run standards button takes you to the run standards dialog box. See
page
89.
Grab sample button takes you to the grab sample load bottles dialog box. See
sample, page
106.
Excursion mode button takes you to the excursion mode dialog box. See
mode, page
110.
Unload bottles button takes you to the unload/replace bottles dialog box. See
Figure 78, page
97.
Figure
67.
Figure 67 Bottle mode dialog box
Run standards,
Grab
Excursion