6.10.1 Excursion with validation
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1. Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See
Figure
92. From left:
•
Bottle 1 = blank
•
Bottle 2 = empty
•
Bottle 3 = 500 ppb sucrose
•
Bottle 4 = excursion bottle
Figure 92 Excursion mode with validation, load bottles dialog box
•
Bottle positions graphics show you the order in which to install the bottles.
•
Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.
•
Back icon takes you to the excursion mode dialog box.
•
Forward icon takes you to the excursion mode setup dialog box.
•
Cancel icon returns you to the previous dialog box. All changes made in any of
the dialog boxes are lost.
2. Close the bottle bay door.
3. Touch the forward icon. The excursion mode setup dialog box appears. See
Figure
93.
Bottle Mode
Page 111