5.0 Defibrillator Energy Test

Equipment Needed:
Power Supply 12V 10A, Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO
Disposable Battery (P/N 1008-002003-01),
AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF),
Defibrillator Analyzer (e.g. QED-6 or Impulse 4000) or equivalent.
Setup:
Connect the HV Connector Cable of the AED PLUS Quick-Test Test Fixture (P/N
9100-0567-TF) to the Defibrillator Analyzer.
Set the AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to Defib Mode.
Set the Power Supply to 12.00V.
Note: Test values below are based of the default adult energy settings (120, 150,
Test
5.1
5.2
5.3
5.4
5.5
5.6
9650-0309-01 Rev. E
and 200 Joules) into a 50 ohm load.
Action
Disconnect the Simulator Cable
from the device, and then turn
the Power Supply on.
Re-connect the Simulator Cable
to the device.
Press the On/Off button and
release it.
Press Flashing Shock button
st
1
shock energy test (default
value 120 joules)
ZOLL AED Pro Service Manual
Expected Result(s)
Verify that the following
prompts are given: PLUG IN
CABLE, PLUG IN CABLE.
Verify that unit voice prompts
ADULT PAD followed by
UNIT OK before unit shuts off
by itself.
Verify that the following
prompts are given: ADULT
PADS, DON'T TOUCH
PATIENT ANALYZING,
followed by "SHOCK
ADVISED, "DON'T TOUCH
PATIENT, PRESS
FLASHING SHOCK
BUTTON.
Verify that unit prompts
SHOCK DELIVERED.
Verify the Energy delivered into
analyzer is 105 joules to 143
joules.
Verify the unit prompts START
CPR followed by CONTINUE
CPR.
Defibrillator Energy Test
Pass
Fail