Chapter 4: Installation

Step 5: Fill the Gas-Liquid Separator

Refer to Figure 6: The Gas-Liquid Separator, as needed.
1.
Open the door of the Analyzer.
2.
Disconnect the tube from the top barb on the left side of the gas-liquid separator (the barb marked "To
NDIR Filter.") See Figure 6: The Gas-Liquid Separator.
3.
Fill the squeeze bottle with deionized (DI) water or the best-quality water available.
4.
Use the squeeze bottle to fill the gas-liquid separator through the center top barb until the water level is
(as shown in Figure 6).
5.
Reconnect the tube to the barb.
GE Analytical Instruments ©2012
Figure 6: The Gas-Liquid Separator
73 of 238
DLM 68100-06 EN Rev. A