Chapter 2: System Description
5-Stream Configuration
The Analyzer also is available in a configuration that permits monitoring of up to five sample streams in the
Online mode. This configuration provides a vial port and check standard port, described in the previous 2-
Stream Configuration section.
Mounting Stand
A mounting stand is available, which accommodates an Analyzer; its reagents, dilution water, and check
standard carboys; and one sample filter panel. The mounting stand allows the Analyzer to be mounted at
various heights for convenience. For safety, the mounting stand must be bolted to the floor or a wall.
Air Purifier
The air purifier accessory makes it possible for the Analyzer to operate using house air as the compressed gas
source. The air purifier removes carbon dioxide and moisture from the air so that it can be used as carrier and
sparge gas. No consumables are required for its operation, but the house air must be oil-free.
Fail Safe Accessory
If the Fail Safe accessory is selected, sensors are installed in the Analyzer to detect when the flow rate of each
sample stream stops. The Analyzer will not attempt to analyze a stream when its flow has stopped, but will
resume analysis of it when the flow starts again. Sensors also are installed that detect when any of the reagents,
dilution water, or check standard needs to be replenished.
IP56
The optional IP56 configuration protects the Analyzer from water jets and the ingress of dust that would
interfere with operation of the Analyzer.
NEMA 4X Cabinet
The optional NEMA 4x cabinet protects the Analyzer from water jets and the ingress of dust and corrosive gases
that would interfere with operation of the Analyzer.