Chapter 4: Installation
N
:
Binary outputs are isolated from system ground and from each other.
OTE
Binary outputs are implemented by relays which are rated for a maximum voltage of 125 VAC or 110 VDC
@ 1 AMP. The maximum switching capacity (resistive load) is 62.5 VA, 33 Watts.
The System Fault Alarm logic is set up so an un-powered instrument indicates an error status. The
System Fault Alarm relay is energized when the Analyzer power is on. Errors cause this relay to de-
energize until the error is cleared.
The Analyzer Complete Output relay will energize momentarily (pulse) when the analysis is complete for
every measurement.
N
:
Binary outputs are isolated from system ground and from each other.
OTE
Binary outputs are implemented by relays which are rated for a maximum voltage of 125 VAC or 110 VDC
@ 1 AMP.
+24 volts and system ground are provided for local use only and are not intended for connections more
than 15 feet away from the Analyzer.
GE Analytical Instruments ©2012
Table 4: Alarm Outputs (TB4)
Pin
Signal
1
+24V
2
+24V
3
+24V
4
GND
5
GND
6
GND
7
BINOUT1_NO
8
BINOUT1_COM
9
BINOUT1_NC
10
BINOUT2_NO
11
BINOUT2_COM
12
BINOUT2_NC
77 of 238
Function
+24 Volts
System
Ground
User Assigned
Alarm #1
User Assigned
Alarm #2
DLM 68100-06 EN Rev. A