Figure 29: Check Standard Setup Screen - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation
To set check standard measurements (Scheduled)
1.
On the Configure Online screen (Setup tab>Online button>Configure Online button), press the Check
Standard button. The Check Standard screen appears.
2.
Press the Frequency button to set how often to measure the check standard.
3.
Press the Start Time button to set the start time using a 24-hr. clock format (e.g., use 13:00 for 1 p.m.).
4.
Press the Concentration button to set the ppm concentration of the check standard.
5.
Press the Tolerance button to set the acceptable range of concentrations for the measurement of the
standard. For example, if the Tolerance is set at 20%, then the acceptable concentration is 20% below
or 20% above the Concentration value.
6.
Press the Check Std. Setup button. The following buttons appear on the Check Standard screen.
GE Analytical Instruments ©2013

Figure 29: Check Standard Setup Screen

Figure 30: Check Standard (with Setup Buttons) Screen
119 of 238
DLM 68100-06 EN Rev. A