Figure 2 Mounting Stand - Clearances - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 4: Installation
44.1 in. (112.0 cm.)
12.0 in. (30.5 cm.)
35.5 in. (90.2 cm.)
Mounting brackets for
filter panel (optional)
71.3 in. (181.3 cm.)
37.21 in. (94.5 cm.)
28.38 in. (72.1 cm.)
Figure 2: Mounting Stand - Clearances
At least two people are required to move the Mounting Stand to the selected location. Bolt the Mounting Stand
to a wall or the floor before placing the Analyzer on it. The horizontal braces shown in Figure 2 can be adjusted
so that the display of the Analyzer is approximately eye-level, but the spacing between the two horizontal
braces must be 94.5 cm.
To mount a Filter Panel on the Mounting Stand, attach the two Filter Panel brackets 30.5 cm apart, as shown in
Figure 2. Further instructions for installation of the Filter Panel are provided in the GE InnovOx Filter Panel
Operation and Maintenance Manual.