Chapter 8: Maintenance
11. Reconnect the tubing to the correct barbs (Figure 46).
12. Perform other maintenance tasks, or close the Analyzer door and return the Analyzer to normal
operation.

Cleaning the Sparger

The sparger requires periodic cleaning, when residue builds up inside.
Wear eye protection and gloves when performing this task.
To clean the sparger
1.
Shut down the Analyzer via the on-screen menu (Maintenance
2.
Wait for the display to turn off, and then turn off power to the Analyzer using the main power switch on
the side of the Analyzer.
3.
Open the door of the Analyzer.
GE Analytical Instruments ©2013
Figure 46: The Gas-Liquid Separator
Warning
158 of 238

Shut Down).
DLM 68100-06 EN Rev. A