Table 1 Analog Outputs (Tb1) - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 4: Installation
N
:
4-20 mA outputs are isolated from system ground.
OTE
4-20 mA outputs are not individually isolated from each other and, therefore, are not intended to be
connected to multiple systems with differences in ground potential.
4-20 mA outputs may be configured by the user to output NPOC, IC, TC, or TOC ppm levels.
GE Analytical Instruments ©2012
Table 1: Analog Outputs (TB1)
Pin
Signal
1
Analog 1 (+)
2
GND_ISO (-)
3
ANALOG 2 (+)
4
GND_ISO (-)
5
ANALOG 3 (+)
6
GND_ISO (-)
7
ANALOG 4 (+)
8
GND_ISO (-)
9
ANALOG 5 (+)
10
GND_ISO (-)
11
ANALOG 6 (+)
12
GND_ISO (-)
Table 2: Binary Input (TB2)
Pin
Signal
1
BININ1+
2
BININ1-
3
BININ2+
4
BININ2-
5
BININ3+
6
BININ3-
75 of 238
Function
4-20mA Stream #1
4-20mA Stream #2
4-20mA Stream #3
4-20mA Stream #4
4-20mA Stream #5
RESERVED
Function
Start/ Stop
RESERVED
RESERVED
DLM 68100-06 EN Rev. A