Consumables Replacement; Replacing The Pinch Valve Tubing - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 8: Maintenance
12. Position the lubricated syringe under the valve.
N
:
If the washer became free upon removal, before reinstalling the syringe into the valve, ensure
OTE
that the washer is in place at the top of the syringe.
13. Screw the syringe base into the threading in the valve by turning the base to the right.
14. Using a 3/32 Allen wrench, tighten the screw that secures the syringe. The top of the syringe moves
independently of the syringe plunger, so it can be aligned with the screw hole.
15. Close the door to the instrument, and restore power to the instrument.
16. Before returning the Analyzer to normal operation, run the Liquid Reactor Flow Diagnostic to confirm
movement of the newly lubricated syringe.
Also confirm that no leaks are present.

Consumables Replacement

Replacing the Pinch Valve Tubing

Pinch valve tubing in the Analyzer's pinch valves must be replaced periodically to ensure proper operation of
these valves. The tubing should be replaced after one year of operating time for neutral pH samples. If samples
are acidic or basic, the tubing should be replaced every 3 to 6 months. The operating life remaining in the tubing
is shown on the Analyzer's Main screen and on the Consumables screen. For this procedure, the Analyzer
should be powered on. Contact GE Analytical Instruments at 303. 444.2009 or 800.255.2595 for a replacement
kit. Technical support in the United Kingdom is available at 44 (0) 161 864 6800.
To replace the tubing
Wear eye protection and gloves when performing this task.
Refer to Figure 54, as needed.
1.
Open the door of the Analyzer.
2.
Locate the replacement tubing from the Pinch Tubing Replacement Kit. Cut the tubing for the following
valves to the same lengths as the tubing you are replacing:
Two Stream and Five Stream Configuration
Valve V1-SH
Valve V5-SH
Valve V3-R
Valve V6-R
Valve V7-R
GE Analytical Instruments ©2013
V1-MS
V2-MS
V6-MS
V7-MS
V2-SH
V3-SH
Warning
163 of 238
DLM 68100-06 EN Rev. A