Step 6: Install Input/Output, Cables, And Accessories - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 4: Installation

Step 6: Install Input/Output, Cables, and Accessories

N
:
Before installing any wiring inside the Analyzer, put on a grounding strap for ESD protection.
OTE
.
The Analyzer's Input/Output (I/O) board is located in the top of the Analyzer under the access cover. Remove the
cover. There are four terminal blocks on the I/O board, with TB1 located closest to the front of the Analyzer. From
front to rear the other terminal blocks are TB2, TB3, and TB4. Consult the tables in this section for a list of
functions on each terminal block. The output and alarm connections should be installed by a qualified
electrician.
Route the cables to the terminal blocks through the pass-through port, labeled Control, on the left side of the
Analyzer. Remove the pass-through cap. Secure the PVC conduit connector ("strain relief" hub) and washer to
the conduit and the Analyzer in the normal manner. Wire should be 22-12 AWG, rated to 300 Volts. Strip lengths
should be 8-9 mm (0.33 in). Install the ferrite (from the Accessories Kit) on the I/O cable, as shown in Figure 6. To
attach the output connections, first remove the terminal block from the I/O board by firmly grasping the
terminal block and pulling it straight out from the board. Make a note of the orientation of the terminal block
before removing. Completely loosen the screw for each pin you are connecting, insert the wire, tighten the
screw, and then gently pull on each connection to make sure the connection is secure. Gently reinsert the
terminal block into the board socket.
Starting from the front of the Analyzer, the terminal blocks are as follows
TB1 Analog Outputs — The maximum 4-20 mA load is 600 ohms.
TB2 Binary Input (Analysis Start/Stop)
TB3 Alarm Outputs (Alarms 3 and 4, System Fault Alarm, and the Binary Analysis Complete Signal)
— The maximum load for these ports is 30 VDC at 1.0 A or 125 VAC at 0.5 A.
TB4 Alarm Outputs (Alarms 1 and 2) — The maximum load for these ports is 30 VDC at 1.0 A or 125 VAC
at 0.5 A.
GE Analytical Instruments ©2012
74 of 238
DLM 68100-06 EN Rev. A