Chapter 5: Calibration and Verification
Figure 20: Editing a Calibration

Performing a Calibration Verification

Verification parallels the options available under calibration menu choices. Verifications may be performed in
TOC, NPOC, TC, or IC analysis modes. As with calibration, you can save multiple verification protocols in each of
the Analyzer's four ranges, and you can save up to 20 verification protocols in total.
For more detailed information regarding options available under verification menus, consult the relevant
sections describing calibration, keeping in mind that you must press the Verification button rather than the
Calibrate button in the Cal/Ver tab, and for each verification protocol you must specify a calibration protocol for
the measurements.
Table 7 lists the approximate number of repetitions that are possible from a 40-mL vial, completely filled with the
standard solution. If Auto Dilution is employed, more repetitions are possible than are shown in Table 7, but the
maximum number of repetitions that can be selected is 20.