Figure 49 Unscrewing The Syringe Base - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 8: Maintenance
6.
Unscrew the syringe base by turning the base to the left.
7.
Remove the syringe from the Analyzer by pulling it down and away from the valve.
The washer should remain in the valve. If the washer comes free of the valve when you remove the
syringe, be sure to capture the washer for reinstallation. You must place the washer back on the
plunger tip when reinstalling.
GE Analytical Instruments ©2013
Figure 49: Unscrewing the Syringe Base
Figure 50: Removing the Syringe
161 of 238
DLM 68100-06 EN Rev. A