Lubricating The Syringe - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 8: Maintenance

Lubricating the Syringe

Perform this instruction after every three months of operation to ensure that the Analyzer is operating with
adequately lubricated syringes.
To lubricate the syringe
Before you begin, locate the Syringe Lubrication Kit that came with the Analyzer's Accessory Kit. You will also
need to provide a 3/32 in. Allen (hex key) wrench.
1.
Turn off the water supply to the Analyzer. If the Analyzer is taking measurements, press the Stop
Analysis button.
2.
Using the Maintenance screen, shut down the Analyzer.
3.
Turn off the power to the Analyzer, and unplug the power cord or disconnect the main power.
4.
Open the door to the inside of the instrument.
5.
Using a 3/32 Allen wrench, loosen and remove the screw that secures the syringe. Remove any washers
under the block, and retain them for reinstallation.
GE Analytical Instruments ©2013
Figure 48: Locating the Syringe Screw and Washers
160 of 238
:
Washers*
*If these washers are NOT installed
with your Analyzer. ignore this
DLM 68100-06 EN Rev. A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents