Cleaning

Cleaning the Monitor

Cleaning Recommendations

Monitor Exterior

4-4
Maintenance: Cleaning
CAUTIONS
Never pour or spray water or any cleaning solution on the
equipment or permit fluids to run behind switches, into
connectors, into the recorder, into the temperature probe
well, or into any ventilation openings in the equipment.
Do not let fluid "pool" around connection pins.
Never immerse monitor or accessories in any liquid.
Do not attach the monitor or accessories to a patient until
it is thoroughly dry.
WARNING
Failure to follow these cleaning recommendations may melt,
distort, or dull the finish of displays and cases; blur lettering on
labels; embrittle cases and lead to cracks and breakage; or
cause equipment failures. Use of non-approved cleaning
agents is not considered normal wear and repair or
replacement of parts is not covered under warranty.
Disconnect the monitor from AC power before cleaning or disinfecting its
surface. The exterior surfaces of DINAMAP Monitors may be cleaned with a
dampened, lint-free cloth. Wipe off all cleaning solutions with a clean, dry cloth
and let air dry for at least 15 minutes. Use one of the following approved
solutions:
Mild soap (diluted)
Commercial diluted bleach solution or bleach wipe
Commercial diluted ammonia solution
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Never use the following cleaning agents:
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol- or petroleum-based cleaning agents
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
Revision C