4.11 Performing Four Arithmetical Operation
Explanation
4-26
Four Arithmetical Operations Function
The following computation results can be shown on display C. " " is displayed at the
front of the value when the computation results are being displayed.
: A + B
: A − B
: A × B
: A ÷ B
: A ÷ B
2
: A ÷ B
2
and
indicates the displayed value on display A and B. "
shown on display A and B and shows the result on display C.
Note
The meanings of the displayed symbols are as follows:
: + (Addition)
: − (Subtraction)
: × (Multiplication)
: ÷ (Division)
: ^ (Exponent)
When the display A function is indicating the elapsed integration time (TIME), the computed
result indication is no data ( -----).
If the value of the display B function is less than or equal to 0.0001% of the rated range,
computation overflow ( --oF-) is indicated.
Application Example
: Displays the result of display A + display B.
Computation example:
Display A
W1
: Displays the result of display A – display B.
Computation example 1:
Display A
W1
W1
Display B
Display C
W2 or W3
W1 + W2
or W1 + W3
W1
Inverter
W2 or W3
Display B
Display C
W3
W1 – W3
Inverter
W3
" adds the values
Wiring method
Any
Wiring method
Any