IM 760401-01E
Setting the rated integration time
The D/A output of integrated values is 5.0 VFS when rated value is input continuously
for the specified time (integration preset time). This is also true when scaling or Σ is
specified.
Selectable range: 0.00.00 (0 hours 00 minutes 00 seconds) to 10000.00.00 (10000
hours, 00 minutes, 00 seconds).
The initial value is 1.00.00. If the time is set to 0.00.00, the DA output value becomes
0 V.
Note
The displayed values of V (voltage), A (current), W (active power), VA (apparent power),
var (reactive power), Vpk (voltage peak), and Apk (current peak) while the MAX hold
function (see section 4.8) is enabled will be the maximum values (MAX) that are held. The
values for D/A output are also set to the maximum values (MAX) that are held.
The D/A output of each output item is configured so that 5.0 VFS is output when the value
corresponding to the range rating of the voltage, current, or power is applied.
Even when scaling constants for voltage, current, and power are specified, the D/A output
is adjusted so that 5.0 VFS is output when the value corresponding to the range rating is
applied.
Also when Σ is specified for the element with the scaling constant of each element is
different, the D/A output is adjusted so that 5.0 VFS is output when the value
corresponding to the range rating is applied to each element.
If
(computation) is specified, the D/A output is 0 V except for efficiency and
average active power.
9.3 D/A Output (Option)
9-7
9