Table of Contents

Advertisement

3: Element 3 (for WT230 only)
Query
HE? <terminator>
Example
HE1
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
IC/IC?
Sets the integration mode/inquires
about the current setting.
Syntax
IC m <terminator>
"m" indicates one of the following
integration modes.
m= 0: Normal integration mode
1: Continuous integration
mode
Query
IC? <terminator>
Example
IC1
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• Changing the integration mode is not allowed
while integration is in progress; execution
error 13 will occur.
• If continuous integration mode is selected,
make sure that the timer preset time is set to
a value larger than "0".
• If normal integration mode is selected, set the
timer preset time to any desired value.
• While recalling or storing is in progress,
execution error 19 will occur.
IM/IM?
Specifies which causes will be allowed
to generate a status byte/inquires about
the current setting.
Syntax
IM m <terminator>
"m" is assigned as follows (0 ≤ m ≤
15).
m= 1: Computation end
2: Integration end
4: Syntax error
8: OVER
Query
IM? <terminator>
Example
IM15
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• If more than one of these causes is to be
allowed, set "m" to the sum of their individual
"m" values. For instance, if all causes are to
be allowed, set "m" to 15 (=1+2+4+8).
IP
Stops integration.
Syntax
IP <terminator>
Description • If an attempt is made to stop integration when
integration has already been interrupted
(stopped), execution error 45 will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
IR
Resets integration.
Syntax
IR <terminator>
IM 760401-01E
Description • If an attempt is made to reset integration
while integration is in progress, execution
error 45 will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
IS
Starts integration.
Syntax
IS <terminator>
Description • 1If an attempt is made to start integration
when integration is already in progress,
execution error 42 will occur.
• If a voltage or current peak overflow, or
overrange takes place when an attempt is
made to start integration, execution error 46
will occur, and integration will not be started.
• While recalling or storing is in progress,
execution error 19 will occur.
KH/KH?
Sets MAX hold/inquires about the
current setting.
Syntax
KH m <terminator>
"m" indicates MAX hold ON/OFF.
m= 0: OFF
1: ON
Query
KH? <terminator>
Example
KH0
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
• Cannot be changed in harmonic
measurement mode. Execution error 19 will
occur.
KL/KL?
Turns ON/OFF key lock/inquires about
the current setting.
Syntax
KL m <terminator>
"m" indicates whether key lock is
ON or OFF.
m= 0: OFF
1: ON
Query
KL? <terminator>
Example
KL0
KV/KV?, KA/KA?, KW/KW?
Sets the scaling constant/inquires about
the current setting.
KV is used for voltage measurement, KA
for current measurement, and KW for
power measurement.
Syntax
When CM0 is set:
KV n <terminator>
KA n <terminator>
KW n <terminator>
When CM1 is set:
KV m,n <terminator>
KA m,n <terminator>
KW m,n <terminator>
"m" indicates element.
m= 0: All elements (Setting not
allowed during inquiry)
13.1 Commands
13
13-5