14.3 Commands
Data Format/Output Format of Normal and Harmonic Measurement Data <ASCII>
14-32
The data format/output format of normal and harmonic measurement data which is
requested by
MEASure[:NORMal]:VALue?
Data Format of Normal Measurement Data
• All data of the <harmonic measurement function> are output in the <NR3> format.
(Example) 123.45E+00
V,A,W,VA,VAR,PF,DEGR,VHZ,AHZ,VPK,APK,MATH
WH,WHP,WHM,AH,AHP,AHM
• The sign of the mantissa will only be applied in case of negative values. However,
phase lead and lag (in case of phase angle (DEG)) will be shown as follows.
LEAD
+180.0E+00
LAG
-180.0E+00
in phase →
0.0E+00
• In case of overrange or computation over,
(i.e. in case the display shows –oL–, –oF–, PFErr, dEGEr, ErrLo, or ErrHi)
• In case no data is present (i.e. the display shows – – – – –),
output.
• The elasped integration time is output as hours, minutes, seconds in the <NR1>
format.
(Example) 999,59,59
• Only one of the six parameters, VHZ1, VHZ2, VHZ3, AHZ1, AHZ2, or AHZ3, can be
measured for frequency. Data that is not measured is set to no data (9.91E+37).
• VHZΣ, AHZΣ, VPKΣ, and APKΣ always output no data (9.91E+37).
Output Format of Normal Measurement Data
The communication output is set ON by any of the commands starting with
"
MEASure[:NORMal]:ITEM
integration time are output according to the following order of priority. Besides, in
case of recalling normal measurement or integration data, the data number will be
output in <NR1> format as well. Data will be output in the following order
corresponding to each element. However, note that for model 760401 only element 1
is valid, and for model 760502 only element 1, 3 and S are valid.
(
Data number in case of recalling)
0.
1. V1
2. A1
3. W1
4. VA1
5. VAR1
6. PF1
7. DEGR1
8. VHZ1
9. AHZ1
10. WH1
11. WHP1
12. WHM1
13. AH1
14. AHP1
14. AHM1
(elapsed integration time)
16. TIME
Each data is divided by a comma"
or
MEASure:HARMonics:VALue?
mantissa: max. 6 digits + exponent: 2 digits
(max. 5 digits in case of negative value)
(The mantissa will be proceeded by a space)
"9.9E+37"(+∞)
" and the normal measured/computed data or elapsed
V2
'
V3
A2
A3
W2
W3
VA2
VA3
VAR2
VAR3
PF2
PF3
DEGR2
DEGR3
VHZ2
VHZ3
AHZ2
AHZ3
WH2
WH3
WHP2
WHP3
WHM2
WHM3
AH2
AH3
vAHP2
AHP3
AHM2
AHM3
,
" and is ended by the terminator <RMT>.
, is as follows.
mantissa: max. 5 digits +
exponent: 2 digits
will be output.
"9.91E+37" (NAN)
V∑
A∑
W∑
VA∑
VAR∑
PF∑
DEGR∑
VHZ∑
AHZ∑
WH∑
WHP∑
WHM∑
AH∑
AHP∑
AHM∑
will be
IM 760401-01E