IM 760401-01E
Output Example of Normal Measurement Data
• Output example for model 760502 after having sent the following commands.
(Sent)
MEASURE:NORMAL:ITEM:PRESET NORMAL
MEASURE:NORMAL:VALUE?
(Received data)
10.004E+00,10.002E+00,10.003E+00,49.041E+00,49.052E+00,
49.047E+00,429.00E+00,429.02E+00,0.8580E+03
(Data contents)
V1:10.004E+00
A1:49.041E+00
W1:429.00E+00
• Output example for model 760503 where measurement data first have been
stored during integration, and while recalling these data, the following
commands have been sent.
(Sent)
MEASURE:NORMAL:ITEM:PRESET INTEGRATE
MEASURE:NORMAL:VALUE?
(Received data)
10,428.60E+00,428.10E+00,428.80E+00,
1.2855E+03,71.450E+00,71.370E+00,71.490E+00,214.31E+00,
8.2342E+00,8.2354E+00,8.2519E+00,24.721E+00,0,10,0
(Data contents)
Recalled data number: 10
W1:428.60E+00
W2:428.10E+00
WH1:71.450E+00
WH2:71.370E+00
AH1:8.2342E+00
AH2:8.2354E+00
Elapsed integration time: 0 (hours), 10 (minutes), 0 (seconds)
Data Format of Harmonic Measurement
All data will be output in the <NR3> format. (mantissa: max. 5 digits + exponent: 2
digits)
Output Format of Harmonic Measurement
The communication output is set ON by any of the commands starting with
"
MEASure:HARMonics:ITEM
source (
) are output according to the following order of priority. Besides,
SYNChronize
in case of recalling normal measurement or integration data, the data number will be
output in <NR1> format as well.
(
0.
Data number in case of recalling)
1.
Frequency of PLL source (
2.VTHD
3.V
8.PF
9.W
Harmonic measurement data will be output for all applicable elements. To find out to
which element the data correspond, use the
• Frequency of PLL Source (
Outputs the fundamental frequency (
PLL source has been set. The input of the PLL source can be found out using
HARMonics:SYNChronize?
: 1 data
VTHD,ATHD
Outputs the harmonic distortion factor of voltage/current. (for either iEC or CSA). The
used computation method can be found out using the
: 51(or 31) data
V,A,W
Rms values of the 1st to 50(or 30)th order→fundamental measured value (1st
order)→harmonic measurement value (2nd order)→ · · · →harmonic measurement
value (50(or 30)th order)
V3:10.002E+00
A3:49.052E+00
W3:429.02E+00
W3:428.80E+00
WH3:71.490E+00
AH3:8.2519E+00
" and the harmonic measurement data or frequency of PLL
SYNChronize
)
4.VCON
5.ATHD
10.WCON
11.VDEG
HARMonics:ELEMent?
) : 1 data
SYNChronize
VHz/AHz
) of the voltage/current for which the
.
14.3 Commands
V∑:10.003E+00
A∑:49.047E+00
W∑:0.8580E+03
W∑:1.2855E+03
WH∑:214.31E+00
AH∑:24.721E+00
6.A
7.ACON
12.ADEG
command.
command.
HARMonics:THD?
14-33
14