YOKOGAWA WT210 User Manual page 142

Digital power meter
Table of Contents

Advertisement

9.3 D/A Output (Option)
9-6
• dFLt-i (integration measurement values are set as default)
Select this when you want to output integration measurement values. Which items
are output to which channel is described below.
Option
Model
760401
ch1
ch2
ch3
ch4
ch5
Output
ch6
ch7
channel
These
ch8
channels
ch9
cannot be
ch10
set.
ch11
ch12
*1
If either one of the display function, V Hz or A Hz, is turned on, the frequency of the
corresponding function is output. If both display functions are OFF, the voltage frequency
of the element assigned to display C is output.
*2
The number corresponds to input element 1, 2, or 3.
Setting an Original Output Format
The items to be output (output function and element) are set per each output channel.
• Setting the output channel
The number of channels depends on the installed options and can be selected
from the following.
• /DA4: 4 channels
• /DA12: 12 channels
• /CMP: 4 channels
• Setting the output function (corresponds to column A in the procedure)
The output function can be set to any of the following.
V (voltage), A (current), P (active power), VAr (reactive power), VA (apparent
power), PF (power factor), VFrq (voltage frequency
Ph (total Watt-hour Wh), Ah (total Ampere-hour), dEG (phase angle), VP (peak
value of voltage), AP (peak value of current), MATH (computation), Ph+ (positive
watt hour value Wh+), Ph– (negative watt hour value Wh–), Ah+ (positive ampere
*2
hour value
), Ah– (negative ampere hour value
further elements can be set)
*1 If either one of the display function, V Hz or A Hz, is turned on, the frequency of
the corresponding function is output. If both display functions are OFF, the
voltage frequency of the element assigned to display C that was illuminated last
when display function V Hz or A Hz was illuminated is output.
*2 For details concerning the positive value of the ampere hour, refer to page 6-3.
• Setting the element (corresponds to colum B in the procedure)
• WT210 (760401) no such element setting available;
• WT230 (760502) element can be selected from 1, 3, or 4
• WT230 (760503) element can be selected from 1, 2, 3, or 4
The element number 4 represents Σ.
/DA4
/DA12
760502
760503
*2
W
W1
W1
Wh
-
W2
Ah
W3
W3
*1
ΣW
ΣW
Hz
Wh1
Wh1
-
Wh2
Wh3
Wh3
ΣWh
ΣWh
Ah1
Ah1
-
Ah2
Ah3
Ah3
ΣAh
ΣAh
/CMP
760401
760502
*2
*2
W
W1
Wh
Wh1
Ah
Ah1
*1
*1
Hz
Hz
These channels cannot be set.
*1
), AFrq (current frequency
*2
), – – – – – – (D/A output 0 V; no
760503
*2
W1
Wh1
Ah1
*1
Hz
*1
),
IM 760401-01E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt230

Table of Contents