YOKOGAWA WT210 User Manual page 188

Digital power meter
Table of Contents

Advertisement

Explanation
IM 760401-01E
Initializing Setup Parameters
Select "YES" in the confirmation menu of the setup procedure and press the ENTER
key to execute the initialization. The initial settings are as follows.
Item
Display A
Display B
Display C
Number of displayed digits
Display update rate
Line filter
Frequency filter
Measurement synchronization A
source
Measurement range
Measurement mode
Wiring method (only WT230) 1P3W
Hold
MAX hold
PC/CT scaling constant
External sensor scaling
constant
Averaging
Crest factor
MATH computing equation
Frequency
Integration
Harmonic measurement (option)PLL source: V1, harmonic distortion factor computation format:
Storage/recalling
D/A output (option)
Comparator (option)
Data output
GP-IB
Serial
Note
Be careful since measurement data will be lost when initializing. However, measurement
data or setup parameters stored in the internal memory will be kept.
If the setup parameters are initialized by a communication command (RC command or
*RST command), parameters related to GPIB and serial communication will not be
initialized.
When Initializing Setup Parameters at Power-On
If the power switch is turned ON while holding down the ENTER key, the WT210/WT230
powers up using initial settings. Keep holding down the ENTER key until the illumination
of all LEDs turn off (step
of the "Opening Message" on page 3-10). All setup
parameters including communication parameters will be initialized as indicated above.
When the setup parameters are initialized, error codes "Err. 60" and "Err. 78" are shown
as messages to indicate the initialization; they do not indicate a malfunction.
12.2 Initializing Setup Parameters
Initial setting
Display function: V, element: 1
Display function: A, element: 1
Display function: W, element: 1
Hi (5 digits)
0.25 s
OFF
OFF
Auto range
RMS
OFF
OFF
P: 1.000, C: 1.000, F: 1.000
scaling ON/OFF: OFF
50.00 A
Averaging type: exponential, attenuation constant: 8
Averaging ON/OFF: OFF
3
WT210: Voltage crest factor
WT230: Efficiency
VHz
Reset condition, integration mode: manual
Integration preset time: 0 hr, 00 min, 00 s
IEC, element: 1
Harmonic measurement function ON/OFF: OFF
Interval: 0hr 00min 00sec, storage/recalling ON/OFF: OFF
Output items: normal measurement items, rated integration time: 1
hr, 00 min, 00 s
Mode: single, determination function: (V1, A1, P1, PF1)
Limit value: refer to section 9.6, display function ON/OFF: OFF
Communication, item: normal measurement setting
Mode: addressable mode, address: 1, status byte: 15,
delimiter: 0
Mode: normal mode, handshaking mode: 0, format: 0,
Baud rate: 9600, delimiter: 0, status byte: 15
12
12-3