Explanation
IM 760401-01E
Selecting the Integration Mode
The following selections are available. The initial value is nor.
• nor: Select this for manual or standard integration mode. Depending on the
integration timer, this instrument will automatically decide the appropriate mode.
• Cont: Select this for the continuous integration mode.
Setting the Integration Timer
Set the integration time. The setting ranges from 0.00.00 (0 hrs, 00 min, 00 s) to
10000.00.00 (10000 hrs, 00 min, 00 s). The initial value is 0.00.00.
• 0.00.00: When "nor" is selected on the integration menu, the manual integration
mode will become valid. If Cont is selected, an error code is displayed when
integration is started, and the integration is not performed.s
• 0.00.01 to 10000.00.00: The time during which integration will be performed when
in the standard or continuous integration mode. The standard or continuous mode
should be selected at the integration mode menu.
6.2 Setting Integration Mode and Integration Timer
6
6-5