Explanation
IM 760401-01E
Setting the Output Item in case of Normal Measurement
• Selecting the Default Setting
Predefined items will be output by the communication function. The following types
of default settings exist and they depend on the model. For a description of the
output format in addressable mode or talk-only mode, see 10-5 to 10-8. For a
description of the output pattern in 488.2 mode, see the description for the
MEASure[:NORMal]:ITEM:PRESet command (page 14-31).
• Normal default setting: dFLt-n
Consists of V (voltage), A (current), W (active power, the above menu shows P),
frequency and displayed data of display C.
• Integration default setting: dFLt-i
Consists of W (active power, the above menu shows P), Wh (watt hour), Ah
(ampere hour), frequency, and integration time.
• Selecting Original Settings
You can set original output items (output function and element) to each output
channel from ch1 to ch14.
• Setting the channel
Select the output channel for assigning the output item from ch1 to ch14.
• Setting the output function (corresponds to column A in the procedure)
Any of the following function can be selected. The initial value is V.
V (voltage), A (current), P (active power), VAr (reactive power), VA (apparent
power), PF (power factor), VFrq (voltage frequency
*1
frequency
), Ph (total watt hour Wh), Ah (total ampere hour), dEG (phase
angle),VP(peak value of voltage), AP(peak value of current),
MATH(computation), t1 (elapsed integration time), Ph+ (positive watt hour
Wh+), Ph– (negative watt hour Wh–), Ah+ (positive ampere hour
(negative ampere hour
*1 If either one of the display function, V Hz or A Hz, is turned on, the frequency of the
corresponding function is output. If both display functions are OFF, the voltage frequency of
the element assigned to display C is output.
*2 For details concerning the positive value of the ampere hour, refer to page 6-3.
• Setting the element (corresponds to column B in the procedure)
The element setting depends on the model and is as follows. The initial value is
"1."
• WT210 (760401) no such element setting available;
• WT230 (760502) element can be selected from 1, 3, or 4
• WT230 (760503) element can be sleected from 1, 2, 3, or 4
The element number 4 represents Σ.
Setting the Output Item in case of Harmonic Measurement
The setting is carried out in the same way as described in section 9.9, "Outputting to
an External Plotter or External Printer." However, graph printing is not possible for
communication output. Only numerical data is output. For details, refer to page 9-23,
9-24.
For a description of the output format in addressable mode or talk-only mode, see 10-
9 and 10-10. For a description of the output pattern in 488.2 mode, see the
description for the MEASure:HARMonics:ITEM:PRESet command (page 14-29).
10.6 Setting the Output Items
*2
), – – – – (no output)
*1
), AFrq (current
*2
), Ah–
10-15