YOKOGAWA WT210 User Manual page 144

Digital power meter
Table of Contents

Advertisement

9.3 D/A Output (Option)
9-8
Relation between the output item and the D/A output voltage
• Frequency
D/A output
Approx. 7.5 V
5.0 V
2.5 V
0.5 V
0.5 Hz
1 Hz
• Integrated value
D/A output
Approx. 7.0 V
140% of rated value input
5.0 V
0
• Other items
Displayed value Output
140%
Approx. 7.0 V
100%
5.0 V
0%
0 V
–100%
–5.0 V
–140%
Approx. –7.0 V
For PF and deg, points in the range from +5 to +7 V and from –5 to –7 V are not
output. If there is an error, the output will be about ±7.5 V.
If the MATH setting is set to efficiency, the output will be +5 V for 100%
For Vp and Ap, the output will be ±5 V when the value is three times (six times
if the crest factor is set to 6) the range.
10 Hz
100 kHz
1 kHz
100 Hz
10 kHz
Rated input
t
0:
D/A output
Approx. 7.5 V
Approx. 7.0 V
5.0 V
–140
–100
0
–5.0 V
Approx. –7.0 V
Approx. –7.5 V
Displayed value
p
Time
t
0
rated integration time
100
140
Displayed value [%]
IM 760401-01E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt230

Table of Contents