YOKOGAWA WT210 User Manual page 293

Digital power meter
Table of Contents

Advertisement

Error Codes regarding Self Diagnosis
Error Code
Description
60
Data failure of setup parameters backup.
(setup parameters other than communication
parameters are set to default)
61
EEPROM (element 1) failure.
62
EEPROM (element 2) failure.
63
EEPROM (element 3) failure.
64
EEPROM (D/A board) failure.
68
Data file failure
(measurement data, setup parameter file failure)
71, 72, 73
DSP communications failure.
75
Internal DSP error.
78
Backup data failure in setup parameters.
79
ROM checksum error.
80
RAM read/write check error.
81
RAM failure on the harmonics measure board
(for models with the /HRM option).
84
DSP sample clock failure.
90
Incorrect board combination.
91
Incorrect board combination.
* Redundancy is provided for EPROMs that handle errors 61, 62, and 63. If one of the errors 61, 62, or 63 appears when the
instrument is turned ON, turn OFF the instrument and turn it back ON. If no errors occur on the second attempt, servicing is
not necessary.
IM 760401-01E
15.4 Error Codes and Corrective Actions
Corrective Action
Service is required.
Service is required
Service is required.
Service is required.
File will be initialized
automatically.
Service is required.
Service is required.
Service is required.
Service is required.
Service is required.
Service is required.
Service is required.
Service is required.
15
15-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wt230

Table of Contents