13.1 Commands
EB/EB?
Sets the element for display B/inquires
about the current setting.
Syntax
EB m <terminator>
"m" indicates element.
m= 1: Element 1
2: Element 2 (for WT230 model
760503 only)
3: Element 3 (for WT230 only)
4: ∑ (for WT230 only)
Query
EB? <terminator>
Example
EB1
Description Parameter error 12 will occur if "m" is set to an
illegal value.
EC/EC?
Sets the element for display C/inquires
about the current setting.
Syntax
EC m <terminator>
"m" indicates element.
m= 1: Element 1
2: Element 2 (for WT230 model
760503 only)
3: Element 3 (for WT230 only)
4: ∑ (for WT230 only)
Query
EC? <terminator>
Example
EC1
Description ∑ Parameter error 12 will occur if "m" is set to
an illegal value.
E,ST,<interface message GET>
Generates a trigger.
Syntax
E <terminator>
ST <terminator>
<interface message GET>
Description This command is valid only during hold mode.
FL/FL?
Determines whether or not frequency
filter is used/inquires about the current
setting.
Syntax
FL m <terminator>
"m" indicates whether frequency
filter is ON or OFF.
m= 0: OFF
1: ON
Query
FL? <terminator>
Example
FL1
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• Filter cannot be switched ON or OFF while
integration is in progress; error 13 will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
H/H?
Determines whether or not to add a
head to measured data output via
communication/inquires about the
current setting.
Syntax
H m <terminator>
"m" indicates whether a header is
added or not.
m= 0: No header added
13-4
1: Header added
Query
H? <terminator>
Example
H0
Description Parameter error 12 will occur if "m" is set to an
illegal value.
HA/HA?
Determines whether or not to turn ON
the harmonic measurement function/
inquires about the current setting.
Syntax
HA m <terminator>
"m" indicates whether the harmonic
measurement function or normal
measurement function is set.
m= 0: Normal measurement
1: Harmonic measurement
Query
HA? <terminator>
Example
HA1
Description • Parameter error 12 will occur if "m" is set to
an illegal value.
• When integration is in progress or being
aborted, harmonic measurement cannot be
performed; error 13 will occur.
• Integration cannot be started when the
harmonic measurement function is in
progress; error 16 will occur.
• While recalling or storing is in progress,
execution error 19 will occur.
• If you switch the harmonic measurement
mode ON/OFF using this command and
query the measurement mode using the "OD"
command immediately afterwards, the
measured data of the previous measurement
mode may be output. To retrieve the
measured data in the new measurement
mode, a wait of approximately 2 seconds is
required after issuing this command.
HD/HD?
Determines whether or not output data
(display, communications, etc.) should
be updated/inquires about the current
setting.
Syntax
HD m <terminator>
"m" indicates the sampling mode.
m= 0: Updates the data at each
sampling rate.
1: Hold
Query
HD? <terminator>
Example
HD0
Description Parameter error 12 will occur if "m" is set to an
illegal value.
HE/HE?
Determines the element of the harmonic
measurement function/inquires about
the current setting.
Syntax
HE m <terminator>
"m" indicates the element of the
harmonic measurement function.
m= 1: Element 1
2: Element 2 (for WT230 model
760503 only)