Setting the Averaging Sample Number/Attenuation Constant
The following selections are available. The initial value is "8".
8, 16, 32, or 64
Setting Averaging ON/OFF
Select the averaging menu once again after having set the averaging values. The
initial value is oFF.
• on: Selecting on and pressing the ENTER key will start averaging and the AVG
indicator will light.
• oFF: Selecting oFF and pressing the ENTER key will stop averaging and the AVG
indicator will extinguish.
Note
The average coefficient is common to exponential average and moving average. If you
change the averaging process, change the averaging coefficient also.
The averaging function that can be used during harmonic measurement is exponential
averaging. The attenuation constant is fixed to 8. Therefore, the setup procedure explained
in this section is valid only during normal measurement.
The averaging function turns OFF when integration is started. It does not turn back ON even
if integration is stopped and reset.
4.7 Using the Averaging Function
4
4-19