Make sure not to greatly vary the input signal when using the comparator
function. Depending on the input signal used for determination, the instrument
may display error codes (i.e. overrange) and this will change the output relays
as follows. When using the output relay as a control signal, make sure to match
these control signals with other equipments to eliminate erroneuous control.
Displayed error
oL (over range)
oF (over flow)
dEGEr (phase angle error)
PFErr (power factor error)
ErrLo (frequency error)
ErrHi (frequency error)
FrqErr (frequency error in case of harmonic measurement) The NC contact is closed.
—— (error when no data are present)
9.4 Comparator Function (Option)
CAUTION
Relay status
The NC contact is closed.
The NC contact is closed.
The NC contact is closed.
The NC contact is closed.
The NC contact is closed.
The NO contact is closed for this
case only.
The NC contact is closed.