8.2 Storing/Recalling Setup Parameters
Explanation
8-6
Storing Setup Parameters
• Stores the current setup parameters which consist of the following. Four
destinations (FiLE1/FilE2/FiLE3/FiLE4) are available.
Measurement range, measurement mode, measurement synchronization source,
scaling settings, averaging settings, input filter settings, MAX hold ON/OFF setting,
computation settings, number of displayed digits, display update rate setting, crest
factor, integration settings, harmonic settings, plotter output settings, store/recall
settings, and communication settings.
• When data are saved in a file and you want to save data in the same file, display C
will show "SAVEd." Pressing the ENTER key will result in overwriting the
previously saved data.
• Setup parameters are saved in another internal memory than measured data.
• Saved setup parameters are backed up by the lithium battery in the same way as
measured data. For a description of the lithium battery life, see section 15.6.
Recalling Setup Parameters
When setup parameters are being retrieved, all setup parameters are being set
accordingly. After that, measurements can be carried out.