6.4 Precautions Regarding Use of Integrator Function
Backup During Power Failures
Operating Restrictions during Integration
6-10
• If there is a power failure while integration is in progress, the integrated value and
integration elapsed time will be backed up. When the power is restored, the display
will show the integrated results up to the time the power failure occurred.
• To start integration after the power is restored, it is necessary to reset integration first.
Certain key operations are restricted during integration, and are shown below.
(START Indicator)
(STOP Indicator)
Function
Wiring system (only WT230)
Measurement synchronization source
Measurement mode
Filter
Measurement range
Crest factor
Scaling
Averaging
MAX hold
Display function
Input element (only WT230)
Number of displayed digits
Display update rate
Hold
Trigger
Integration mode
Integration timer
Integration start
Integration stop
Integration reset
Harmonic measurement (option)
Store/recall
Comparator
Plotter, printer
Zero-level compensation
Ο
: Settings can be changed
×
: Settings cannot be changed. Attempts will result in an error code.
⋅ When integration is started during auto range, the measurement range will change to fixed range.
⋅ If you start integration when averaging is ON, averaging is turned OFF. It does not turn back ON
even if integration is stopped and reset.
Integration status
Integration
Integration in
reset
progress
Not lit
Lit
Not lit
Not lit
Ο
Ο
×
Ο
×
Ο
×
Ο
×
Ο
×
Ο
×
Ο
×
Ο
Ο
Ο
Ο
Ο
Ο
Ο
Ο
Ο
×
Ο
Ο
Ο
Ο
Ο
Ο
Settings cannot be changed,
but can be displayed
Ο
Settings cannot be changed,
but can be displayed
×
Ο
×
Ο
×
Ο
×
Ο
×
Ο
Store possible
Ο
Ο
Ο
Ο
×
Ο
Integration
interrupted
Not lit
Lit
Ο
×
×
×
×
×
×
×
Ο
Ο
Ο
Ο
×
Ο
Ο
Ο
×
Ο
×
×
Store possible