Maintenance, cleaning and disinfection
INFORMATION
When the joystick is activated while the brake is deacti-
vated, the control unit emits an error signal on the control
panel. If this is not the case, there is a malfunction that
must be corrected immediately by a specialist dealer.
9.2
Changing the fuse
The fuse (for values see section „Technical Data" > Table
„Electrical equipment") is located in a fuse holder at the rear
end of the battery case (see Fig. 51).
1. Open the cap of the fuse holder.
2. Pull out the fuse.
3. Insert a new fuse into the holder. Make sure that the
fuse is pressed into the centre of the spring contacts
and that it is not positioned at an angle.
4. Close the cap until you feel it snap back into place.
222
Ottobock
|
Figure 51
Fuse holder
1
Cover open
2
Fuse, inserted
9.3
Wheel replacement
NOTICE
Risk of damage due to uncontrolled movements. When
raising the power wheelchair, e.g. with a jack, secure it
against slipping and tipping sideways by using a suitable
block under the drive unit sustainer.
1
2
C1000 DS