Malfunctions / troubleshooting
LED indicator
Seat function LEDs (see Fig. 40, item 5)
Backrest LED is
Back angle adjustment
flashing
motor fault
Seat bottom LED
Seat tilt motor fault
is flashing
2 LEDs below
Seat height adjustment
seat bottom are
motor fault
flashing
Backrest / seat
Electric seat function
bottom LED is
temperature warning
flashing
Table 11
Attendant control error messages
218
Ottobock
|
Warning/Error
Faulty cabling or plug contact;
Defective actuator
Overheating due to excessive load
Cause
Check cabling / plug connec-
tions;
Contact specialist dealer
Cool down phase
Possible corrective action