Table 3-13: D distance
Omega IV Compact
Omega V Long
Omega V non motorized Long
Innova IQ Table
NOTE:
* in case of a Innova IQ Table upgrade, there is the possibility of a distance of 1595 mm
(62.8 in).
NOTE:
Refer to table
Section 2, Mounting Requirements

3.3.5 Water Pipe Requirements

An Innova LC System uses a Performix 160 A X–Ray tube with a recirculating chiller.
Two water hoses are supplied to allow water circulation between Innova LC Positioner and
chiller.
In some countries, it is forbidden to run electrical cables and water pipes in the same conduit.
In this case, two separate conduits are required. But then a problem arises at the level of the
Innova LC Positioner entrance.
Depending on the method used, one of the two options shown hereafter
3-7
and
Illustration
Chapter 3 Physical Requirements of Room
Due to the plastic bushing used in the USA to protect cables from the sharp edges
of conduits it is necessary to place the cable conduit inside the table cable access
opening but the height of the outcoming conduit plus bushing is limited to 1/2 in
(12.7 mm).
Chemical anchors Pull out efforts and recommendations in
3-8) must be used.
ANGIO
NA
1395 mm (54.9 in)
1278 mm (50.3 in)
NA
1278 mm (50.3 in)
NA
1278 mm (50.3 in)
NA
for pull out effort on each fixation bolts.
Direction 5160944-13-1EN, Revision 2
CARDIO
NA
1395 mm (54,9 in)
1395 mm (54,9 in)
1395 mm (54,9 in)*
(IllustrationsIllustration
NEURO
Chapter 5,
71