Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
3 Structural Requirements
3.1 Room Size
For Room size dimensions, refer to
For additional details, refer to
3.2 Door Size Requirements
Minimum door sizes also apply to hallways and elevators. For additional details, refer to
7, Section 1, Shipping
3.2.1 Door Height
The minimum door height (to accommodate Innova positioner on its dolly) is 1.980 m (78 in). If the
height is limited to 2 m (79 in), you will need a fall over cabinet.
3.2.2 Door Width
The minimum door width needed (to accommodate the Innova LC shipping dolly) is:
•
1.165 m (46 in) with protective side rail,
•
1.096 m (43 in) with one protective side rail removed on site.
NOTE:
Door widths are based on a
corridor. Calculations need to be made for accommodation of equipment through
narrower corridors.
3.3 Floor
3.3.1 General Vascular GEMS Policy
GEMS's Customer is responsible for the structural analysis and mounting of the base plates. If
GEMS is forced to mount the base plate, the LCT must hire a structural engineer to design and
approve the mounting method and provide GEMS with an engineering report.
The floor level cannot exceed a general levelness of 5 mm (0.2 in) for any 2 meters (79 in).
66
Chapter 4, Section 1, Room Layout
Information.
The floor slabs on which the equipment is to be installed must have a levelness of
1 mm (0.04 in) per meter (40 in). Position of baseplates and table basement
depends on the type of installation. The two types of installation are given in
Illustration
3-2.
Chapter 4, Section 2, Room Layout
straight–in approach requiring a 2.44 m (96 in) wide
Direction 5160944-13-1EN, Revision 2
Drawings.
Considerations.
3 Structural Requirements
Chapter