7.4 CLS300 digital switching point setup

7.4
CLS300 digital switching point setup
Use the following procedure to set up the switch point for the CLS200 Digital model.
1. Set the 0% Application Range Setting (Mode 19):
– Install the device in the process.
– Use the M button to navigate to Mode 19.
– Set up the process to provide the recommended condition for your process from the
– Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
– Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
– Set the sensor value to 95% of the value obtained in step 1d (0.95 x counts). Press ⇑
– After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate
– Press M to exit
2. Set the 100% Application Range setting (Mode 20):
– Install the device in the process.
– Use the M button to navigate to Mode 20.
– Set up the process to provide the recommended condition for your process from the
56
table below:
Application
Material
General
dry solids
low viscosity liquids
Demanding
hygroscopic / wet solids
high viscosity and high
conductivity liquids
Interface detec-
liquid A / liquid B
tion
foam / liquid
Record the count value.
to edit the value of each digit, and press ⇓ to move the cursor to the next digit.
the procedure and save the new value. The display now shows the new value for the
0% Application Range setting.
table below:
Application
Material
General
dry solids
low viscosity liquids
Demanding
hygroscopic / wet solids
high viscosity and high conductivity
liquids
Interface detec-
liquid A / liquid B
tion
foam / liquid
Setup conditions
Material in tank should be at the lowest
point it could reach during the normal pro-
cess.
Material in tank should be at the lowest
point it could reach during the normal pro-
cess. The sensor should have the minimum
possible material build up.
Material in tank should be at the lowest
point it could reach during the normal pro-
cess.
Setup conditions
Material in tank should be at
the highest point it could
reach during the normal pro-
cess.
Material in tank should be at
the highest point it could
reach during the normal pro-
cess.
Sensor immersed in the mate-
rial with the highest dielectric
constant
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA