Technical data
11.1 Pointek CLS200/300
11.1.4
Alarm outputs
solid-state switch
time delay
hysteresis
failsafe operation
terminal
Counts The frequency value is always represented in 'counts': see Counts (Page 88).
1)
11.1.5
Diagnostics
input
11.1.6
Mechanical
Electrode - CLS200 digital
Model
Length (max)
Rod
5500 mm/
216.5"
Sani-
5500 mm/
tary
216.5"
Cable
30000 mm/
1181.1"
Or 1.4404 material.
1)
82
galvanically isolated non-polarity sensitive transistor rated 30V DC or peak AC
max., 82mA max voltage drop below 1 V typical @ 50 mA
With intrinsically safe: barrier required
•
for ATEX: U
•
for FM/CSA: FM/CSA connection drawing (Page 115)
controlled by software 2 delay timers: alarm ON delay and alarm OFF delay)
100% adjustable, in 1 count increments
Failsafe High or Failsafe Low
removable terminal block, 2.5 mm
Process Connections
•
Threaded: ¾", 1", or 1 ½" BSPT
(R), BSPP (G) or NPT; 1 ¼" NPT
only
•
Welded flange: ASME: 1", 1
½", 2", 3" or 4" DN 25, 40, 50,
80 or 100
1", 1 ½ ", 2", 2 ½" and 3" sanitary
clamp
•
Threaded: ¾", 1", or 1 ½" BSPT
(R), BSPP (G) or NPT; 1 ¼" NPT
only
•
Welded flange: ASME: 1", 1
½", 2", 3" or 4" DN 25, 40, 50,
80 or 100
= 30V I
= 200mA P
=350mW C
i
i
i
1)
max.
2
reed contact: for test function
Extension
Tensile
(max)
316L
stain-
n/a
1)
less steel
316L
stain-
n/a
1)
less steel
FEP (Fluori-
180 kg/
nated Eth-
400 lbs
ylene
Polymer)
Operating Instructions, 05/2021, A5E47904060-AA
=0 L
=0
i
i
on the display
Wetted Parts
•
316L
stainless steel (optional
1)
PFA coating)
•
FKM seals (optional FFKM)
•
PPS probe (optional PVDF)
•
316L
stainless steel (optional
1)
PFA coating)
•
FKM seals (optional FFKM)
•
PPS probe (optional PVDF)
•
316L
stainless steel (optional
1)
PFA coating)
•
FEP jacketed cable
•
FKM seals (optional FFKM)
•
PPS probe (optional PVDF)
Pointek CLS200/CLS300 digital