Technical reference
A.7 Sensor test
① Active measurement section
② Active shield
③ Pointek CLS300
④ Empty tank
The relative lengths of the measurement section and Active Shield section can be specified to
suit a particular application. If the measured range will be short relative to the total length of
the electrode, specify a short measurement section. This increases the achievable resolution
of the measurement, since any change in level will be greater relative to the length of the
measurement section.
The powder-coated aluminum enclosure provides reliable operation in environments with
dust, moisture, and high-frequency interference.
A.7
Sensor test
You can initialize a sensor test by activating the reed contact with a magnet
the housing. For details, see Magnet-activated sensor test in Mode 2: Sensor test (default:
OFF) (Page 42).
A test magnet will be supplied with the instrument.
1)
90
⑤
Full tank
⑥
Buffer
⑦
Frequency (f) ≈ K / capacitance (C)
Operating Instructions, 05/2021, A5E47904060-AA
from outside
1)
Pointek CLS200/CLS300 digital