Wiring; Electrical Connection - Siemens Pointek CLS200 Operating Instructions Manual

Hide thumbs Also See for Pointek CLS200:
Table of Contents

Advertisement

Wiring

6.1

Electrical connection

• Observe the specifications of the examination certificate valid in your country.
• Observe the laws and regulations valid in your country for electrical installations in
• Refer to Hazardous area installation (Page 104) if applicable.
• Ensure that the available power supply complies with the power supply specified on the
• Shipping plugs in the cable inlets must be replaced by suitable screw type glands or
• For CE installations, use a cable with a braided metallic shield (or armoured cable where
• The lid must not be opened in wet locations while the unit is powered. (A wet location is
• Cable entry devices and closing elements of unused apertures must meet a temperature
Pointek CLS200/CLS300 digital
Operating Instructions, 05/2021, A5E47904060-AA
WARNING
potentially explosive atmospheres.
product nameplate and specified in the examination certificate valid in your country.
dummy plugs, which are appropriately certified for transmitters with explosion-proof
protection.
applicable).
a location where water or another conductive fluid may be present and is likely to
increase the risk of electric shock.)
range from min. -40 to 10 K above maximum ambient temperature.